Four sondes deployed distant to the 6200 dcp – YSI 6200 User Manual

Page 52

Advertising
background image

YSI 6200 DAS USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543 Page 5-7

Battery
(charged Solar or AC)

6200 DCP

Radio Transceiver

6-Series Sondes

Water Temperature
Conductivity
Dissolved Oxygen
pH
ORP
Depth/Level
and more...

Radio
Antenna

Field Station

Junction Box

MS-8
Patch Cable

User-provided
SDI-12 cable

Sonde #1

Sonde #2

Breakout Box

Sonde #3

Sonde #4

Breakout Box

User-provided
SDI-12 cable

Figure 5.5 4-Sonde Connection Remote to the 6200 DCP

Four Sondes Deployed Distant to the 6200 DCP

If more than two sondes are required and are remote to the 6200 DCP installation, you will need
two 6502 Breakout Boxes, a 6507 Patch Cable, and a 6508 Junction Box. The breakout boxes
should be mounted near the deployed sondes. The junction box is mounted near the 6200 DCP,
and user supplied cable connects the two breakout boxes to the junction box. The patch cable
connects the junction box to the 6200 DCP. A maximum of 250 ft. (75 m) is recommended
between the farthest sonde and 6200 DCP. See Figure 5.5 for an illustration.

Note that there are 4 different sonde addresses assigned in this connection scenario. SDI-12
addresses 1, 2, 3, and 4 would be logical, but any unique addresses will work. The wiring
parallels what is described in the previous description and the pin out and color designations in
Figure 5.7 are appropriate for the wiring. You can see from this diagram that to go beyond 4
sondes you will need an additional junction boxes and breakout boxes.