2.7 Testing Sensor Function

Now that you are able to view readings, you may wish to perform some preliminary tests of
sensor function. As you have seen some readings display 0.000 in the checkout testing. For
example, rain fall and accumulation, wind speed and direction and solar radiation probably have
zero readings. Here are some quick tests you can perform.

For wind speed/direction locate the plastic arrow protruding from the MET Suite head assembly.
Rotate the MET head assembly on the support so that the arrow points north (0 degrees). East
would be 90, south 180 and west 270 degrees. Direction will not be displayed unless there is
measurable wind speed. To check wind speed set up a small fan to provide continuous air
motion in front of the propeller. During interrogation of the 6200 DCP, EcoWatch DCP will
upload readings showing wind speed and direction. In the example below the wind speed is
about 5.4 mph and the direction is 0

°

, which is due north.

For rain rate/accumulation testing, remove the top assembly of the rain gauge to locate the
tipping buckets. Carefully tip the buckets from one position to the other 5 to 10 times in