Two sondes deployed near the 6200 dcp – YSI 6200 User Manual

Page 50

Advertising
background image

YSI 6200 DAS USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543 Page 5-5

Battery
(charged Solar or AC)

6200 DCP

Radio Transceiver

Radio
Base Station

Computer with
EcoWatch DCP

6-Series Sondes

Water Temperature
Conductivity
Dissolved Oxygen
pH
ORP
Depth/Level
and more...

Radio
Antenna

Base Station

Field Station

Sonde #1

Sonde #2

Breakout Box

MS-8
Patch Cable

Figure 5.3 2-Sonde Connection Near to the 6200 DCP

Two Sondes Deployed Near the 6200 DCP

If two sondes are required and are near to the 6200 DCP installation, then the 6504 Breakout Box
with a 6507 Patch Cable provide all you need to connect these sondes. The breakout box must be
mounted within 6 ft (2 m) of the 6200 DCP and the sondes close enough that their cables can
reach the breakout box. Standard cables can be ordered up to 200 ft. (60 m). The breakout box
has two MS-8 connectors for weatherproof connections. The patch cable is connected to the
breakout box using a feed-through gland. The “flying leads” attach to the terminal connector
provided with the breakout box. Refer to Figure 5.5 for the SDI-12 pin out.