YSI 6200 User Manual

Page 19

Advertising
background image

YSI 6200 DAS USER Manual

YSI/Massachusetts 508.748.0366, Fax 508.748.2543 Page 2-10

3. From Main menu press 4-Report, to enter the menu that allows you to assign the parameters
and the units of your choice. In the example we select five parameters...one temperature, one
conductivity, one pH and two for DO. After assigning parameters press 0 to return to Main.

4. From Main menu press 6-Advanced, then 2-Setup. Verify that the screen below matches
what you see. It is very important that the Auto sleep options are activated, especially when the
sonde is being set up for deployment. For more details see Section 3 and your sonde manual.

Press 0’s to return to the statement that asks you to press Y or N to exit the sonde menu. Press Y
to exit the sonde menu. Close the Sonde-COM1 window and disconnect the adapters and cables.
Now you should connect the 6-series sonde to the 6200 DCP.

-----------------Main-----------------

1-Run

4-Report

2-Calibrate

5-Sensor

3-System

6-Advanced

Select Option (0 for previous menu):

------------Sensors enabled-------------

1-(*)Temperature

5-(*)ISE1 pH

2-(*)Conductivity

6-( )ISE2 NONE

3-(*)Dissolved Oxy

7-( )ISE3 NONE

4-( )Pressure-Abs

Select Option (0 for previous menu):

-----------------Main-----------------

1-Run

4-Report

2-Calibrate

5-Sensor

3-System

6-Advanced

Select Option (0 for previous menu):

--------------Report setup-------------

1-(*)Temp C

A-( )Resist Ohm*cm

2-( )Temp F

B-( )TDS g/L

3-( )Temp K

C-( )TDS Kg/L

4-( )SpCond mS/cm

D-( )Sal ppt

5-(*)SpCond uS/cm

E-(*)DOsat %

6-( )Cond mS/cm

F-(*)DO mg/L

7-( )Cond uS/cm

G-( )DOchrg

8-( )Resist MOhm*cm H-(*)pH

9-( )Resist KOhm*cm I-( )pH mV

Select Option (0 for previous menu):

----------------Advanced---------------

1-Cal Constants

3-Sensor

2-Setup

4-Data Filter

Select Option (0 for previous menu):

-------------Advanced setup------------
1-(*)VT emulation

2-( )Power up to Menu

3-( )Power up to Run

4-( )Comma Radix
5-(*)Auto sleep RS232

6-(*)Auto sleep SDI12

7-( )Multi SDI12

Select Option (0 for previous menu):

Advertising