Chapter 2: prerequisites – Labconco Protector XL Floor-Mounted (Walk-In) Laboratory Hoods 98606 User Manual

Page 13

Advertising
background image

9

Product Service 1-800-522-7658

Chapter 2: Prerequisites


W/ Rear Lower Vertical Sash Closed and Front

Vertical Sash Open or Horizontal Doors Open

Hood Size

Standard Model Description

Face

Velocity

Exhaust

Volume

Static

Pressure

Loss

Sashes

Open

4' Vertical

48" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

550 CFM

0.10" H

2

O

100 fpm

680 CFM

0.16" H

2

O

1

5' Vertical

60" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

720 CFM

0.16" H

2

O

100 fpm

900 CFM

0.24" H

2

O

1

6' Vertical

72" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

900 CFM

0.18" H

2

O

100 fpm

1120 CFM

0.28" H

2

O

1

8' Vertical

96" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

1250 CFM

0.12" H

2

O

with Vertical-Rising Sashes

100 fpm

1560 CFM

0.19" H

2

O

1

8' Horizontal

96" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

1760 CFM

0.19" H

2

O

with horizontal-sliding sashes

100 fpm

2200 CFM

0.30" H

2

O

2

10' Horizontal

120" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

1090 CFM

0.08" H

2

O

with horizontal-sliding sashes

100 fpm

1360 CFM

0.12" H

2

O

1

12' Horizontal

144" Protector XL Floor-Mounted (Walk-In) Hood

80 fpm

1330 CFM

0.10" H

2

O

with horizontal-sliding sashes

100 fpm

1660 CFM

0.15" H

2

O

1

16' Horizontal

192" Protector XL Floor-Mounted (Walk-In) Hood

with horizontal-sliding sashes

100 fpm

2260 CFM

0.10" H

2

O

1

NOTE: 8' Floor-Mounted Hoods are offered with both
vertical-rising and horizontal-sliding sashes, and have
different CFM’s and pressures.

Proper blower selection can be determined from these
exhaust requirements and the total system static
pressure loss. Contact Labconco Customer Service at
800-821-5525 for assistance in sizing a blower system.