Connecting to the protector hood exhaust system – Labconco Protector XL Floor-Mounted (Walk-In) Laboratory Hoods 98606 User Manual

Page 30

Advertising
background image

26

Product Service 1-800-522-7658

Chapter 3: Getting Started

Connecting to the Protector
Hood Exhaust System







The exhaust connection should be
installed by a qualified HVAC
contractor
.

The exhaust connections on your hood
has been designed for 12" nominal pipe
(12.75" OD) to allow for minimum static
pressure loss with proper transport
velocities away from the hood. Consult
Labconco Customer Service should you
require help sizing your blower for the
exhaust volume and total system static
pressure loss.

The selected exhaust duct material
should match the hood procedures and
chemicals used to ensure compatibility.

Connecting the Electrical
Supply Source to the
Protector Fume Hood

Prior to connecting any electrical wiring to the fume
hood structure, refer to the hood identification plate for
the proper electrical requirements of your specific
model.






!

WARNING: The weight of the exhaust
ductwork system must be supported
independently of the hood superstructure. Do
not allow this weight to be supported by the
hood structure as damage to the hood may
occur.

WARNING: The building electrical supply system for
Protector Hoods should include overload protection.
A switch or circuit breaker should be in close
proximity to the equipment and within easy reach of
the operator. The switch or circuit breaker is to be
marked as the disconnecting device for the equipment.
Consult the NEC-2002 for proper installation.

!

!