Refer to the following table if your fume hood fails to operate
properly. If the suggested corrective actions do not solve
your problem, contact Labconco for additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Remote blower and
lights won’t operate

Wires not connected
at junction boxes or
switches.

Check connection of switches.

Check connection to control box on top
of unit.

Circuit

breakers

tripped in building
electrical supply.

Reset circuit breakers.

Remote blower
won’t operate, but
lights work

Blower wiring is
disconnected.
Belt broken.

Inspect blower wiring and switch.

Replace belt.

Blower

motor

is

defective.

Replace blower motor.

Fume hood blower
operates but lights
will not operate

Lamp not installed
correctly.

Inspect lamp installation.

Lamp is defective.

Replace lamp.

Lamp

circuit

breaker

in building is tripped.

Reset the lamp circuit breaker.