Chapter 5: Using Your Protector Fume Hood

Operating the Horizontal-
Sliding Sashes

Horizontal-sliding sashes allow the operator the
flexibility of arranging the sashes at a position suitable
for various procedures. The glass sashes have a 68.15"
sash opening providing a large viewing area. The
airflow requirements are sized for the 50% open sash
condition for the 8' model and 25% for the 10', 12', and
16' models.

Operating the Blower

Your Protector Fume Hood utilizes a remote style
blower, which can be activated by turning the blower
switch to “ON.” You can validate the hood
performance by watching smoke drawn into the hood
face opening.

Operating the Lights

Your Protector Fume Hood utilizes a factory-wired
fluorescent light to illuminate the hood interior. Simply
turn the light switch to “ON” to operate.

Working in your Protector
Fume Hood

Planning

• Thoroughly understand procedures and

equipment required before beginning work.

• Arrange for minimal disruptions, such as room

traffic or entry into the room while the hood is
in use.


Start-up

• Turn on fluorescent light and hood blower.

• Slowly raise the sash or slide open one sash

panel.

• Check the baffle air slots for obstructions.

• Allow the hood to operate unobstructed for 5

minutes.

6'

8'

6'

8'

16'

12'

10'

8'

5'

4'

10' 12' 16'

5'

4'

10' 12' 16'