Distillation grids – field installation – Labconco Protector XL Floor-Mounted (Walk-In) Laboratory Hoods 98606 User Manual
Page 53
49
Product Service 1-800-522-7658
Chapter 7: Modifying Your Protector Fume Hood
Installing Guardian™ Digital
Airflow Monitor or Guardian™
Jr. Airflow Monitor
The Guardian Digital Airflow Monitor P/N 9743211
continuously monitors face velocity through the fume
hood opening. The Guardian Jr. Airflow Monitor P/N
9743202 continuously monitors airflow through the
exhaust. The fume hood right corner post is factory
prepared to mount either monitor. Contact Labconco
Customer Service to order.
Figure
7-5
Figure
7-6
Distillation Grids – Field
Installation
The distillation grids have been strategically placed
with the vertical rod centerlines in front of the lower
baffle and middle baffle. The distillation grids allow
the hood user to mount motors, stirrers, and other
apparatus.