Routine service operations, Your next step – Labconco Protector XL Floor-Mounted (Walk-In) Laboratory Hoods 98606 User Manual
Page 49
![background image](/manuals/738686/49/background.png)
45
Product Service 1-800-522-7658
Chapter 6: Maintaining Your Protector Fume Hood
Routine Service Operations
Front Panel Removal:
1. Lift the front panel up and then away from the hood
to provide access to the top.
Changing the Fluorescent Lamp:
1. Turn light switch to “OFF.”
2. Remove the front panel as noted earlier.
3. Reach over the front header of the hood and remove
the knock out plugs at both ends of fixture. Lift
fixture up.
4. Remove the fluorescent lamp by pushing it out of
the spring-loaded lamp socket and swinging it out
of the other lamp socket.
5. Install the new lamp by reversing the removal
procedure.
Your Next Step
After you understand the maintenance procedures, you
are ready to proceed to Chapter 7: Modifying Your
Protector XL Floor-Mounted Fume Hood.