DATAMAN 848 MANUAL

Operation

41

exist in different sockets and the operator corrects one of
the sockets, the timer resets and waits the specified time
for each succeeding error. Time-out intervals can be set
from 1 to 999 seconds. The default value is 5 seconds.

Insertion Test

This option performs the device-insertion check of the
chips in the sockets. The insertion check includes poor pin
contact, pin count mismatch (the pin count of the chip
designated in the software does not match the pin count of
the actual chip in the socket), device in wrong position,
device upside-down, short-circuit between pins, and chip
damage. Results are displayed at each socket's LED

Device ID Check

This option performs a device signature and manufacturer
match test. With the chip selected and plugged into a
socket, DATAMAN-848 checks the device ID and displays
the results of each check on the LED display.

Verify Passes

Checking this option will instruct the DATAMAN-848 to
perform device verification with the buffer data when
programming is complete. When verify passes is enabled,
one of the three verify options (as described below) must
be set.

Verifying Options

The following three options are available for verification of
data retention following programming: verify twice with 5%
V

CC

, verify twice with 10% V

CC

, and verify once with V

CC

.

These options will only be enabled if the 'Verify passes'
option (see above) is enabled.

(·) Twice V

CC

± 5 %