7 limigard controls, 8 sil mode – standard sil (optional), Figure 4.31 – inputs – Flowserve MX Electronic Actuator User Manual

Page 77

Advertising
background image

77

Limitorque MX Electronic Actuator FCD LMENIM2306-06 – 10/13

flowserve.com

Figure 4.31 – Inputs

4.19.7 Limigard Controls

The Limigard controls is available for the MX that presents the user with redundant inputs at Inputs a and 5 for close,
Inputs 1 and 4 for open, and Inputs 2 and 3 for stop. Both inputs need to be asserted in order for the actuator to move
open or close. Only 1 stop input needs to be asserted in order for the actuator to stop.

NOTE: These three inputs are normally configurable, however, when this custom mode is enabled, they are set to the
predefined configuration and cannot be changed by the customer.

Input a (normal default - ESO) terminal 30:
Set as remote close
Input 1 (normal default · Open Inhibit) terminal 34:
• Set as remote open
Input 2 (normal default - Close Inhibit) terminal 35:
• Set as remote stop
Input 3 (normal default - Stop) terminal 26:
Input 4 (normal default - Open) terminal 25:
Input 5 (normal default - Close) terminal 27:

4.19.8 SIL Mode – Standard SIL (Optional)

Custom software is available for the MX that permits the User to establish certain performance characteristics for
partial stroke (PS) testing and ESD.

NOTE: These three inputs are normally configurable. When this SIL mode is enabled, input 0 is preconfigured to
ESD, input 1 is pre-configured to partial stroke, and both inputs cannot be changed by the customer. Input 2 can be
configured to user input, inhibit, ESD or CSE.

Input 0 (normal default - ESD) terminal 30:
• Set as enabled, ESD, signal present
Input 1 (normal default - Open Inhibit) terminal 34:
• Set as enabled, partial stroke function, signal present = active
Input 2 (normal default - Close Inhibit) terminal 35:
• Set as disabled, user input, signal absent = active

When this mode is selected, the ESD overrides are set so that ESD had priority over stop, local, and network ESD. You
cannot change that Input 0 is ESD, but it can be enabled/disabled, set to signal present/absent, or you can change the
ESD action.

NOTE: These three inputs are kept as open-stop-close inputs. When the custom mode is enabled, the wire mode is
preset to 4-wire control and cannot be altered.

Input 3 (normal default - stop) terminal 26
Input 4 (normal default - open) terminal 25
Input 5 (normal default - close) terminal 27

CHANGE

INPUTS

MODE

4-WIRE OK?

MULTI POSITION

2 OK?

CHANGE

INPUT-0?

STATUS

(ON)-OK?

M1 MOVE TO

(XXX%) OPEN?

PROP BAND

(XXX%) OPEN

DEAD BAND

(XXX%) OPEN

(OFF)

NOTE 1: If using 1 Multi Position then Input 0 will be set as your M1 Move To and

Inputs 1 and 2 are standard configurable. If 2 Multi Positions are used then Input 1

is set as your M2 Position and input 2 is standard configureable.

INPUT 0: ON, M1 Move To, signal present

INPUT 1: OFF, User Input, signal present

INPUT 2: OFF, User Input, signal present

INPUT 3: ON, STOP

INPUT 4: ON, OPEN

INPUT 5: ON, CLOSE

SEE NOTE 1

3 WIRE

MAINT

3 WIRE INCH

Advertising
This manual is related to the following products: