5 modes of operation – Fluke Biomedical 8000 Victoreen User Manual

Page 22

Advertising
background image

Victoreen 8000
Operators Manual

2-10

Wait a moment while the 8000 analyzes the exposure then displays the measured kVp, exposure, time
and mAs (mAs will be zero because the mAs input was not used).

80.0

kVp Avg

100

msec

79.2

kV Eff

392

mR

81.1

kV Peak

0.0

mAs

RADIO

75%

LOW

+ 10 ms

Please

MODE %kV

SENS

DELAY

Wait…


The 8000 will immediately display Please Wait… in the lower right corner of its screen while it prepares to
take another exposure.

After the 8000 beeps and displays Make Exposure in the lower right corner of the display, another
exposure may be made.

To stop making exposures and return to the main menu screen, press the MODE button.

2.5 Modes of Operation

The NERO mAx Nero has ten modes of operation as listed below (in selection order):

1. RADIO

UNIT

2. MAMMO

SETUP

CAL

3. FLUORO

HVL

EXP

4. AMSE

CT EXP

1Ø PULSE

5. CT

EXP

AMSE ZERO

6. EXP

FLUORO

75%

MAMMO 80%

7. HVL

LOW

RADIO

90%

8. CAL

HIGH

10 mS

MODE %kV

9. SETUP

SENS DELAY

10. UNIT ID


The NERO mAx's mode of operation is controlled from the main screen. When the menu cursor (blinking
reverse video) is moved to a mode selection, by pressing the mode key or using the up and down arrow
keys, all of the options available for that mode are displayed. Pressing a softkey (1 thru 4) moves the
menu cursor into the selected menu column and starts it blinking. Pressing the softkey again will move
the blinking menu cursor up through the column of available options. Pressing the up and down arrow
keys moves the blinking menu cursor up or down through the selected column of available options. After
options have been highlighted, the enter key is pressed to start the selected mode with the selected
options. Pressing the mode key at any time returns to the mode select screen.

NERO mAx Operating Modes

Radio Mode

Radio mode is used to make measurements on tungsten target, aluminum filtered radiographic x-ray
generators. Radio mode simultaneously measures kVp, exposure, exposure time and mAs from a single
radiographic exposure. Radiographic mode may also be used to verify the NERO mAx’s kV calibration.