Fluke Biomedical 8000 Victoreen User Manual

Page 56

Advertising
background image

Victoreen 8000
Operators Manual

2-44

Exposure and rate measurements made with external ion chambers are calculated using the stored
calibration factor (R/nC or Gy/nC) for the selected external chamber and air density correction. For more
information about entry of air density information (temperature and pressure), see SETUP mode.


2.5.9 Setup Mode

The Setup Mode is used to setup various features of the NERO mAx. From the setup screen the user
can do the following:

• Set the instruments internal real time clock.
• Set the temperature and pressure used in air density correction of exposure measurements.
• Select exposure units of either Roentgens or Grays.
• Turn automatic printing on or off.
• Turn the clock display on or off and select normal or reverse video on screen clock display.

To Set the Time

SETUP

CLOCK

>

OFF

CAL

PRINT

>

OFF

HVL

UNITS

>

R

EXP

AIR

−−− >

20.5 C

734 mmHg

CT EXP

DATE

−− >

Oct. 10,

1996

AMSE

TIME

−−>

10

30

45

MODE SELECT

HOURS

MINUTES

SECONDS

Using the mode key, go to the SETUP screen and use the select softkey (1) to select TIME. When time is
selected, the time selection blinks and a highlight (reverse video) will extend across the other 3 display
fields. The hours, minutes and seconds keys select and increment hours, minutes or seconds. Holding
down any of these keys (hours, mins, secs) will continuously increment the selected unit. Once a unit of
time is selected, it may also be incremented or decremented with the up or down keys. Once the desired
time has been set, press the ENTER key to accept it.

The NERO mAx keeps time in a military or 24 hour format. As a result, the time must be entered in the
same 24-hour format. For example 9:30 AM is entered as 9:30 but 9:30 PM must be entered as 21:30.

To Set the Date

SETUP

CLOCK

>

OFF

CAL

PRINT

>

OFF

HVL

UNITS

>

R

EXP

AIR

−−− >

20.5 C

734 mmHg

CT EXP

DATE

−− >

Oct.

10,

1996

AMSE

TIME

−−>

10 30 45

MODE SELECT

MONTH

DAY YEAR

From the SETUP screen, use the select softkey (1) to select DATE. When DATE is selected, the date
selection blinks and a highlight (reverse video) extends across the other 3 display fields. The month, day
and year keys select and increment months, days or years. Holding down any of these keys (month, day,
year) will continuously increment the selected unit. Once a unit is selected, it may be incremented or
decremented with the up or down keys. Once the desired date has been set, press the ENTER key to
accept it.