Victoreen 8000
Operators Manual

2-26


WARNING

Extreme caution should be used when making
connections to the mAs terminals of the X-ray
generator or detector. Improper connections may
result in injury, damage to the NERO mAx, and/or
damage to the x-ray generator.

If results are to be printed, plug the printer cable into the printer port at the rear of the NERO mAx then
plug the other end of the printer cable into the printer and turn the printer on (see Section1.5--Printing).
Plug the AC adapter into the NERO mAx if needed (see Section 1.4--Battery Operation and Charging)
and turn the instrument on.

EXP

CT EXP

AMSE

FLUORO

MAMMO

LOW

RADIO HIGH

MODE SENS

Select AMSE mode and sensitivity, then press enter.

Please wait…

FILTER = 50 - 100

AMSE LOW

MODE SENS

Please wait while the NERO mAx prepares to take an exposure. Note that the selected filter kV range is
displayed. If the NERO mAx detects a fault condition such as an invalid filter card or position, an error
message is displayed and the user must correct the error to continue (see Section 4.3--Error Messages--
for more information). When the NERO mAx is ready to take an exposure it will beep and prompt for an
exposure.

MAKE EXPOSURE

FILTER = 50 – 100

AMSE LOW

MODE SENS

Start the exposure.