Victoreen 8000
Operators Manual

4-14

4.6 Waveforms - Partial kV Waveforms

Exposure Results, Low Sensitivity

kVp Avg

kV Eff

kV Peak

mSec

mR

40.5 39.7 42 10.6

2















kV Waveform

Low Sensitivity

0

5

10

15

20

25

30

35

40

45

0

2

4

6

8

10

12

mSec

kV

kV

NOTE


In the above kV waveform, no data is plotted below
approximately 39 kV. This is because there is
insufficient data available to fully construct the kV
waveform. This occurs when exposures are made
near the detectability limit of the NERO mAx. To
remedy this situation, switch to high sensitivity or to
the next lower filter range. Below is the kV
waveform from the same exposure made using
high sensitivity.

Exposure Results, High Sensitivity

kVp Avg

kV Eff

kV Peak

mSec

mR