Victoreen 8000
Operators Manual


5-2

5. Disconnect both battery leads (red and black) from the battery.

6. Cut the two tie wraps that retain the battery.

7. With a flat bladed screwdriver, pry the battery up from its mounting plate and remove the battery.


NOTE

The battery is held to the mounting plate with
double-sided tape.

8. Replace the battery, in its original position at the rear of the battery mounting plate with its positive

lead to the right. Use the double-sided tape from the old battery to hold the new battery in place
and secure the battery to the mounting plate with two new tie wraps. See Appendix F--
Replacement Parts--for part numbers.

9. Reconnect the battery leads and reassemble the NERO mAx in the reverse order of disassembly.

10. Turn the NERO mAx on and verify that the NERO mAx powers up successfully. If the NERO mAx

does not power up properly, please call Fluke Biomedical at 440.248.9300.

5.3 Routine Cleaning

CAUTION


Do not immerse the Model 8000. The unit is not
waterproof. Liquid could damage the circuits. The unit
should be kept clean and free from dirt and
contamination. The unit may be cleaned by wiping with
a damp cloth or using any commercially available
cleaning or decontaminating agent.