Fluke Biomedical 8000 Victoreen User Manual

Page 50

Advertising
background image

Victoreen 8000
Operators Manual

2-38

9.8

R/min

HVL INTERN

RATE LOW

MODE CHAMBER

MODE

SENS

SAVE

After the exposure has started, the NERO mAx displays the exposure rate. When the rate has stabilized
or the desired rate has been achieved, press the SAVE softkey (softkey 4) to save the rate measurement.

9.8

R/min

HVL INTERN

RATE LOW

TURN OFF

MODE CHAMBER

MODE

SENS

BEAM

Turn off the x-ray beam. The NERO mAx has stored the measured exposure rate and will set up for the
next exposure as soon as the beam turns off.

0.0

R/min

Please Wait…

HVL INTERN

RATE LOW

MODE CHAMBER

MODE

SENS

REJECT

While the exposure rate from the first exposure is displayed, the NERO mAx prepares to take the second
exposure. Please wait while the NERO mAx prepares to take the next exposure. If the NERO mAx
detects a fault condition such as a high offset, an error message is displayed and the user must correct
the error to continue (see Section 4.3--Error Messages--for more information).

9.8

R/min

MAKE EXPOSURE # 02

HVL INTERN

RATE LOW

MODE CHAMBER

MODE

SENS

REJECT

The NERO mAx is now ready to take the next exposure. Position an aluminum absorber in the beam and
make the exposure. If the previously measured exposure is not acceptable, press the reject softkey to
retake the exposure. If an overrange occurs, an error message is displayed and the exposure must be
taken over.