3 specifications – Fluke Biomedical 8000 Victoreen User Manual

Page 8

Advertising
background image

Victoreen 8000
Operators Manual

1-4

1.3 Specifications

Kilovoltage

Measured during the first 480 ms of exposure

Accuracy:

0.5 kV or ± 1%

Reproducibility:

0.5 kV or ± 1%

Range:

W/Al

30 - 60 kV

50 - 100 kV

80 - 160 kV

Mo/Mo

22 - 35 kV

Mo/Rh

22 - 40 kV

Mo/Al

22 - 49 kV

Rh/Rh

25 - 49 kV

Rh/Al

25 - 49 kV

Time To Display

Radio & Mammo: 3 seconds for 0.1 second exposure

1 second for each 32 ms of exposure time

Fluoro & AMSE: 15 seconds for all exposures

Time

Measured during entire exposure at 90% rise/fall of waveform.

Accuracy: 1

ms

Range:

1 ms to 60 sec

Exposure/Exposure & Rate

Measured during entire exposure; kVp corrected.

Accuracy: ±

5%

Reproducibility:

(Radio & Mammo Modes) ± 2% or 2 mR

Range:

1 mR minimum

mAs and mA

Measured during entire exposure

Accuracy: ±

2%

Reproducibility:

± 1% or 0.2 mAs

Range:

1 - 1000 mA

HVL

Accuracy: ±

5%

Range:

.1 - 99.9 mmAl

Physical

Display:

240 x 60 pixel, super twist LCD w/ccfl backlight

Power:

115 or 230 VAC External Supply. Rechargeable internal batteries supply more than
4 hours of continuous service with overnight charge.