Victoreen 8000
Operators Manual

2-24

318

μR/pulse

0

mAs/pulse

0.28

R/min

FLUORO PULSED LOW

MODE TYPE SENS

SAVE

After the exposure starts, the NERO mAx displays exposure rate (R/min, mR/pulse) and mAs/pulse.
When the desired exposure rate and/or mAs/pulse are displayed, press the SAVE key to acquire kV data.

318

μR/pulse

0

mAs/pulse

0.28

R/min

FLUORO PULSED LOW

TURN OFF

MODE TYPE SENS

BEAM

After kV data is acquired, the NERO mAx prompts to turn off the beam.

318

μR/pulse

0

mAs/pulse

0.28

R/min

FLUORO PULSED LOW

ANALYZE

MODE TYPE SENS

DATA

After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.

318

μR/pulse

0

mAs/pulse

0.28

R/min

FLUORO PULSED LOW

Please

MODE TYPE SENS

Wait…

After data analysis is complete, kV, exposure rate and mAs/pulse are displayed. If an overrange is
detected an error message is displayed. If the calculated kV is above or below the selected filter kV
range, “High” or “Low” is displayed instead of kV (see Section 4.3--Error Messages--for more information).
After the exposure data is displayed, the NERO mAx displays a "Please wait..." message as it prepares
for the next exposure.