Troubleshooting waveforms-overshoot 4-13 – Fluke Biomedical 8000 Victoreen User Manual

Page 75

Advertising
background image

4

Troubleshooting

Waveforms-Overshoot

4-13

NOTE


The calculated exposure time for this 100
millisecond exposure, taken with a ten millisecond
delay, is now correct and the overshoot at the
beginning of the exposure has been skipped. The
waveform is now only 90 milliseconds in length
because the NERO mAx delayed data acquisition
for ten milliseconds.

Exposure Results with 10 Millisecond Delay

kVp Avg

kV Eff

kV Peak

mSec

mR

70.3 69.8 71.8 101.5

227

kV Waveform

10 Millisecond Delay

0

10

20

30

40

50

60

70

80

90

0

10

20

30

40

50

60

70

80

90

100

mSec

kV

kV

Radiation Waveform

10 Millisecond Delay

0

2000

4000

6000

8000

10000

12000

0

10

20

30

40

50

60

70

80

90

100

mSec

counts

Channel A

Channel B