Operation

Modes of Operation

2-29

Plug the AC adapter into the NERO mAx if needed (see Section 1.4--Battery Operation and Charging)
and turn the instrument on.


WARNING

An electric shock hazard exists between the ion
chamber bias connector and ground.

EXP

CT EXP

AMSE

FLUORO

MAMMO LOW

RADIO

HIGH

10 mm

MODE SENS BEAM

Select CT EXP mode, high or low sensitivity and beam width (1-10 mm), then press the ENTER key.
Pressing the BEAM softkey (2) or the UP key increments the beam width by 1 mm, pressing the DOWN
button decrements the beam width by 1 mm.

Please wait…

CT EXP

HIGH

10 mm

MODE SENS BEAM

Please wait while the NERO mAx prepares to take an exposure. If the NERO mAx detects a fault
condition such as a high offset, an error message is displayed and the user must correct the error to
continue (see Section 4.3--Error Messages--for more information).

0.00

R

CT EXP

HIGH

10 mm

MODE SENS BEAM

RESET

When the NERO mAx is ready to take an exposure it will beep and display zero exposure.

6.72

R

CT EXP

HIGH

10 mm

MODE SENS BEAM

RESET

After the exposure ends, the measured exposure for the scan is displayed. If an overrange occurs, an
error message will be displayed (see Section 4.3--Error Messages--for more information). When the
exposure ends, the NERO mAx auto-resets before the next exposure while retaining the measured
exposure data on screen. The NERO mAx beeps when auto reset is complete and the NERO mAx is
ready for the next exposure. Press the RESET softkey (4) to zero the display or press the mode key to
exit. Do not press the RESET softkey when the beam is on. Inaccurate readings will result.