Victoreen 8000 operators manual – Fluke Biomedical 8000 Victoreen User Manual

Page 80

Advertising
background image

Victoreen 8000
Operators Manual

4-18

Radiation Waveform

-20 Millisecond Delay

0

1000

2000

3000

4000

5000

6000

0

20

40

60

80

100

120

140

160

180

mSec

counts

Channel A

Channel B



















The NERO mAx Excel Add-In may be used to view kV and radiation waveforms to determine the optimum
measurement delay to use. Information on using the NERO mAx Excel Add-In to view waveforms may be
found in the NERO mAx Toolkit for Excel Instruction Manual. In addition, a digital storage oscilloscope
may be connected to the scope output on the rear panel of the NERO mAx readout to view radiation
output waveforms. Information on using the NERO mAx scope output may be found in Section 1.6 of the
NERO mAx instruction manual.