Fluke Biomedical 8000 Victoreen User Manual

Page 33

Advertising
background image

Operation

Modes of Operation

2

2-21

MAKE EXPOSURE

FILTER = 50 - 100

FLUORO CONT

HIGH

MODE TYPE SENS

Start the fluoroscopic exposure.

4.49

R/min

4.1

mA

FLUORO CONT

HIGH

MODE TYPE SENS

SAVE

After the exposure starts, the NERO mAx displays exposure rate and mA. When the desired exposure
rate and/or mA are displayed, press the SAVE key to acquire kV data.

4.49

R/min

4.1

mA

FLUORO CONT

HIGH

TURN OFF

MODE TYPE SENS

BEAM

After kV data is acquired, the NERO mAx prompts to turn off the beam.

4.49

R/min

4.1

mA

FLUORO CONT

HIGH

ANALYZE

MODE TYPE SENS

DATA


After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.

89.6

kVp Avg

87.8

kV Eff

4.34

R/min

91.9

kV Peak

4.1

mA

FLUORO CONT

HIGH

Please

MODE TYPE SENS

Wait…

After data analysis is complete, kV, exposure rate and mA are displayed. If an overrange is detected an
error message is displayed. If the calculated kV is above or below the selected filter kV range, “High” or