Operation

Modes of Operation

2

2-27

250.9

mR/frame

45.6

mAs/frame

AMSE LOW

MODE SENS

SAVE

After the exposure starts, the NERO mAx displays exposure/frame and mAs/frame. When the desired
exposure rate is displayed, press the SAVE key to acquire kV data.

250.9

mR/frame

45.6

mAs/frame

AMSE LOW

TURN OFF

MODE SENS

BEAM

After kV data is acquired, the NERO mAx prompts to turn off the beam.

250.9

mR/frame

45.6

mAs/frame

AMSE LOW

ANALYZE

MODE SENS

DATA

After the beam is turned off, the NERO mAx analyzes the fluoro exposure kV data.

65.6

kVp Avg

57

msec/frame

65.0

kV Eff

250.9

mR/frame

67.2

kV Peak

45.6

mAs/frame

AMSE LOW

Please

MODE SENS

Wait…

After data analysis is complete, kV, exposure rate, mAs/frame and time/frame are displayed. If an
overrange is detected an error message is displayed. If the calculated kV is above or below the selected
filter kV range, “High” or “Low” is displayed instead of kV (see Section 4.3--Error Messages--for more
information). After the exposure data is displayed, the NERO mAx displays a "Please wait..." message as
it prepares for the next exposure.