Optional – Ivoclar Vivadent IPS e.max CAD Abutment Solutions User Manual

Page 29

Advertising
background image

29

Combination firing (Crystallization and Stain/Glaze firing in one step)

Please observe the following procedure for the combination firing:
– Extrude the ready-to-use IPS e.max CAD Crystall./Glaze Paste from the syringe and mix.
– If a slight thinning is desired, the ready-to-use glaze may be mixed with a small amount of IPS e.max CAD Crystall./Glaze

Liquid.

– Apply IPS e.max CAD Crystall./Glaze Paste evenly on the areas to be glazed using a small brush. Avoid to apply too thick

a glaze layer. Avoid "pooling", especially on the occlusal surface of the abutment crown.

– Too thin a glaze layer may lead to an unsatisfactory gloss.
– Apply characterizations with IPS e.max CAD Crystall./Shades and/or IPS e.max CAD Crystall./Stains. For that purpose,

extrude the Shades and Stains from the syringe and mix. If necessary, slightly thin them using IPS e.max CAD Crystall./
Glaze Liquid. However, the consistency should still remain pasty.

– Apply mixed Shades and Stains directly into the unfired glaze layer using a fine brush. More intensive shades are

achieved by several staining procedures and repeated firing, not by applying thicker layers.

– To imitate the incisal area and translucency of the hybrid abutment crown in the incisal and occlusal third, IPS e.max

CAD Crystall./Shades Incisal may be used. The cusps and fissures can be individualized using Stains.

After glazing and staining, the Combination firing is conducted in a compatible ceramic furnace
(e.g. Programat

®

CS or Programat P500). When placing the objects into the furnace and setting the firing parameters,

observe the following points:
– Place the restoration in the centre of the IPS e.max CAD Crystallization Tray.
– A maximum of 6 units can be positioned on the firing tray and crystallized in the Combination firing with IPS e.max CAD

Crystall./Glaze Paste.

– Remove restoration from the furnace after completion of the firing cycle (wait for the acoustic signal of the furnace).
– Allow the objects to cool to room temperature in a place protected from draft.
– Do not touch the hot objects with metal tongs.

Important:

– The glazing material must neither reach the bonding surface to the Ti base nor the screw channel, as this may

compromise the accuracy of fit. Check the interface before firing and carefully remove any contamination.

– On the abutment, do not apply any materials to the bonding surface to the crown, as this might compromise the fit

of the crown.

Completing the IPS e.max CAD Ceramic Structure

– Staining T

echnique on the Blue Restoration

Optional:

For minor shape adjustments (e.g. proximal or occlusal contact points), IPS e.max CAD Crystall./Add-
On is available. The detailed procedure is described on page 33.

Important

– Conduct the Combination firing on the IPS e.max CAD Crystallization Tray using the stipulated firing

parameters.

Observe the firing parameters for IPS e.max CAD MO and IPS e.max CAD LT. Firing parameters
see page 64

Note:

If a restoration made of IPS e.max CAD MO and one made of IPS e.max CAD LT are to be crystallized in the same

firing, the firing parameters for IPS e.max CAD MO must be used!