Crystallization and firing parameters – Ivoclar Vivadent IPS e.max CAD Abutment Solutions User Manual

Page 65

Advertising
background image

65

Crystallization and Firing Parameters


The following points should be observed for ceramic furnaces used for the crystallization of IPS e.max CAD:
– Crystallization should be carried out in an Ivoclar Vivadent ceramic furnace (e.g. Programat CS, Programat P300, P500,

or P700).

– If you use other, untested ceramic furnaces, please consult Ivoclar Vivadent about their compatibility with IPS e.max

CAD.

Basically, the following applies:
Ceramic furnaces without
– function for controlled (long-term) cooling
– vacuum function
cannot be used.
– Before the first crystallization and every six months after that, the ceramic furnace must be calibrated.
– Depending on the mode of operation, more frequent calibrations may be required. Observe the instructions of the man-

ufacturer.

The following aspects should be observed for conducting the crystallization:
– Use only IPS Object Fix Putty or Flow as an auxiliary firing paste to place the restoration directly on the IPS e.max CAD

Crystallization Tray.

– IPS e.max CAD restorations must not be directly placed on the IPS e.max CAD Crystallization Tray and the Pins, i.e. with-

out auxiliary firing paste, for crystallization.

– Use only the IPS e.max CAD Crystallization Tray and the corresponding Pins, since they store the heat necessary for slow

and above all tension-free cooling of the glass-ceramic.

– Always conduct the crystallization under vacuum.
– Remove IPS e.max CAD objects from the furnace after completion of the firing cycle (wait for the acoustic signal of the

furnace).

– Allow the objects to cool to room temperature in a place protected from draft.
– Do not touch the hot objects with metal tongs.
– Do not blast or quench the objects.

crystallization and firing p arameters

please fold out the page

Advertising