For internal operation, the 9530/8530 contains a timer and mode generator which generates an
internal

T

0

clock that is used to trigger all the channel timers. System modes are controlled via the

Mode menu.

Using Continuous Mode

The Run/Stop button starts and stops a continuous pulse stream at the rate specified by the Rate
menu. This corresponds to the normal output mode for most pulse generators. To generate a
continuous stream of pulses:

Within the System Mode menu select Continuous mode.

Within the Rate menu:

o Select the system oscillator or external clock in frequency.
o Set the desired frequency

Pressing the

Run/Stop

key will now generate a stream of

T

0

pulses at a rate specified by the period

parameter.

Using Single Shot Mode

To generate a single pulse with every press of the Run/Stop key:

Within the System Mode menu select Single Shot mode.

Pressing the

Run/Stop

key will now generate one pulse.

Using System Burst Mode Function

The Run/Stop button generates a stream of ‘N’ T

0

pulses, where the ‘N’ is specified by the Burst

parameter. The rate is specified in the Rate menu. Pressing the

Run/Stop

button while in the burst

is in process will stop the output. After the burst has been completed, pressing the

Run/Stop

button

will generate another burst. To generate a burst of pulses set:

Within the System Mode menu:

o Select mode to be Burst
o Set the Burst parameter to produce the number of pulses desired

Using the System Duty Cycle Function

The Run/Stop button starts a continuous stream of T

0

pulses, which oscillates on for ‘N’ pulses and

off for ‘M’ pulses, where ‘N’ and ‘M’ are specified by the On/Off parameters respectively. The rate at
which the pulses are generated is controlled in the Rate menu. To generate a stream of pulses
which will oscillate on for ‘N’ pulses and off for ‘M’ pulses set:

Within the System Mode menu:

o Set the mode to Duty Cycle.
o

Set the On parameter to the number of pulses to produce during the on cycle (‘N’).

o

Set the Off parameter to the number of pulses to suppress during the off cycle (‘M’).

Within the Rate menu:

o Set the Source to either the systems internal oscillator or to external mode.