Scpi, Ommand, Ummary – Quantum Composers 9530 Series User Manual

Page 37

Advertising
background image

9530/8530 Manual Version 4.8

Page 37

9530/8530 SCPI Command Summary

Keyword:

Command:

Sub-
Command:

Parameter
Range:

Notes:

:INSTrument

The units’ upper level command keyword.

:CATalog

?

Returns a comma separated list of the names of all
channels.
Example: A two channel unit would return T

0

, CHA,

CHB

:FULL

?

Returns a comma-separated list of the names of all
the channels and their associated number.
Example: a two channel unit would return T

0

, 0,

CHA, 1, CHB, 2

:COMMands

?

Returns an indentured list of all valid SCPI
commands

:NSELect

0 - 8

Selects a channel using the numeric value

:SELect

T

0

/ CH[A-H]

Selects a channel using the identifier

:STATe

0/1 or OFF/ON

Enables/Disables the selected channel output. If no
channel has been selected the command is applied
to T

0.

If T

0

is selected all outputs are affected.

Enabling T

0

is the same as pressing the RUN

button.


Keyword:

Command:

Sub-
Command:

Parameter
Range:

Notes:

:DISPlay

Command to change the units’ display settings.

:ENABle

0/1 or OFF/ON

Enables/Disables the display, also locks the
keypad. Once the system is rebooted the display
will default to the on state.

:MODe

0/1 or OFF/ON

Command to enable/disable the automatic display
update function. Setting to 1 will force a display
update when a command is received via serial
communications. Setting to 0 will turn this feature
off. The default setting is 0
*Note: Using this feature will slow down the
communication response.

:BRIGhtness

1 - 8

Command to change the display brightness. 8 is
the brightest setting.

:UPDate

?

Query only. Will force the display to be updated
with the current parameters. This command will
only update the display once for every time the
command is sent.