:INFOrmation




?



Query only. Returns model,
serial number, firmware
version, and FPGA version
numbers. The same as the
*IDN? Command.

:NSID

?

Query only. Returns firmware
and FPGA identification
numbers.

:CAPS

0/1 or OFF/ON

The default value is 1, which
means the unit is not case
sensitive. 0 means the
commands sent to the unit
must be capitalized.
*Note: To change this
parameter the unit must be
power cycled before the
command will take effect.

IEEE

488.2 Common Commands

Command:

Parameter Range:

Notes:

*IDN

?

Query only. Returns model, serial number, firmware version, and FPGA
version numbers.

*RCL

1 - 24

RECALL

*RST

RESET. Resets parameters only same as *RCL 0

*SAV

1 - 24

SAVE

*TRG

Creates a soft trigger input.

*TTG

Creates a soft trigger for the trigger input only. Used when the dual trigger
option is enabled.

*GTG

Creates a soft trigger for the gate input only. This command is only active
when the dual trigger option is enabled.

*LBL

?

Used to query the label of the last saved or recalled configuration.

String Value

Command to attach a string label to the current settings. The string must
be in double quotes and no longer than 14 characters. Command must be
followed by a *sav [1/2/n] command to take effect.
*Note: To see the label on the screen a display update or reboot must take
place.

*ARM

0/1 or OFF/ON

Resets all channel counters simultaneously when the channels are in
either single shot or burst mode.
*Note: The system must be in continuous mode (this command is
functionally the same as pressing the Run/Stop button).