Displaying calibration values – HEIDENHAIN TNC 640 (34059x-05) ISO programming User Manual

Page 522

Advertising
background image

Manual operation and setup

15.8 Calibrating a 3-D touch trigger probe

15

522

TNC 640 | User's ManualDIN/ISO Programming | 1/2015

Calibration using a calibration sphere

Proceed as follows for manual calibration using a calibration sphere:

In the

Manual Operation mode, position the ball tip above the

center of the calibration sphere

Select the calibration function: Press the

CAL. R

soft key
Enter the diameter of the sphere
Enter the safety clearance
Enter the start angle
Enter the number of touch points
Select Length measurement, if applicable
Enter Datum for length, if applicable
Start the probing procedure: Press the machine

START button. The 3-D touch probe probes all

required touch points in an automatic probing

routine and calculates the effective ball-tip radius.

If probing from opposite orientations is possible,

the TNC calculates the center offset
Checking the results
Press the

OK soft key for the values to take effect

Press the

END soft key to terminate the calibrating

function. The TNC logs the calibration process in

TCHPRMAN.html.

In order to be able to determine the ball-tip center

misalignment, the TNC needs to be specially

prepared by the machine manufacturer.
Refer to your machine manual.

Displaying calibration values

The TNC saves the effective length and effective radius of the

touch probe in the tool table. The TNC saves the ball-tip center
misalignment in the touch-probe table, in the

CAL_OF1 (principal

axis) and

CAL_OF2 (minor axis) columns. You can display the values

on the screen by pressing the

TOUCH PROBE TABLE soft key.

During calibration, the TNC automatically creates the

TCHPRMAN.html log file to which the calibration values are saved.

Make sure that you have activated the correct tool

number before using the touch probe, regardless

of whether you wish to run the touch probe cycle
in automatic mode or in the

Manual Operation

operating mode.

For more information about the touch probe table,

refer to the User’s Manual for Cycle Programming.

Advertising