ProSoft Technology MVI69-HART User Manual

Page 159

Advertising
background image

MVI69-HART ♦ CompactLogix Platform

Reference

HART Multi-drop Master Communication Module

User Manual

ProSoft Technology, Inc.

Page 159 of 169

March 29, 2012

5.10 Example Field Terminations

Up to 15
instruments per
channel

One instrument per channel

Multi-drop Mode
(address 1 to 15)

Point to Point
Mode
(address 0 to 15)

Point to Point
Mode with
Analog
(address 0 only)

Power Mode A
24VDC connected
to HART power

terminals 1 and 2

on terminal block

Master Powered
instruments on all

4 channels

OK
Install 250Ω 1KΩ

OK
Install 1K resistor

OK

Power Mode B
HART power

terminals 1 and 2

on terminal block

left open

Loop Powered
instruments on 1

or more channels

OK

OK

OK
- differential or

single ended

analog signal

Self Powered
instruments on 1

or more channels

Not possible

OK

OK
-differential or

single ended

analog signal

5.10.1 Notes

1 Both channels on the unit must be either Master Powered (Power Mode A) or

must be Loop or Self powered (Power Mode B). You cannot mix Power Mode

A and Power Mode B on the same unit. You can mix Loop powered and Self

powered instruments on the same unit. However, you cannot mix Loop

powered and Self powered instruments on the same channel.

2 For Master Powered, Multi-drop Mode and Point to Point Mode, a 1K, 1W

resistor must be connected across the + and - input terminals of each HART

channel. Resistors are supplied with each product.

3 When configuring loop powered instruments in a Power Mode B type multi-

drop network, use only one power supply per HART channel if you wish to

maintain channel to channel isolation. If channel to channel isolation is not

required, the same power supply may be used for both channels. However, a

separate series resistor is required for each channel. An appropriately sized

resistor must be connected in series with the power supply for each channel.

The value and power rating of the resistor is dependant upon the number of

instruments in the loop and the length and gauge of the pair of wires

connecting the instruments to the unit. 250 Ohm, 3 Watt is a good, first

choice value.

4 If field instrument wiring is shielded, terminate shields at the ground terminals

of the HART connector. Do not terminate the shields at the instrument.

Advertising