Fluke 5725A User Manual

Page 104

Advertising
background image

5725A

Instruction Manual

4-34

latch up by eliminating positive feedback resulting from positive output voltages from
U100. Op amp U105 is simply a low-noise inverting amplifier that scales the sensor
output to 10 dB below the 5700A Oscillator output level. The output of the calibration
circuit, AC/AC, goes to the 5700A during calibration through the analog monitor
section.

4-56. Analog Monitor Section

The 16-channel multiplexer, U151, takes a variety of inputs from all analog assemblies
and provides a means of switching them to either a comparator, U154A, or to the B-RCL
line via K152. During normal operation, the multiplexer output is routed exclusively to
the comparator. Calibration requires routing the multiplexer output to the B-RCL line,
which is routed back to the 5700A. The B-RCL line is measured by the precision adc
circuit on the 5700A DAC assembly. All inputs to the multiplexer are scaled so that
inputs are within its

±

15V power supply range. Table 4-4 shows the signal name of the

multiplexer input, its description, origin, and purpose.

The state of the multiplexer is controlled by the microprocessor via data latched into
U152.

Table 4-4. Signals Monitored by the Analog Monitor Section

STANDBY
VOLTAGE

LIMITS(1)

SIGNAL

NAME

PHYSICAL

PARAMETER

SOURCE

PURPOSE

-0.125V

+400V IM

Supply Current

Power Supply

Monitoring

+0.125V

-400V IM

Supply Current

Power Supply

Monitoring

0.26V

+400V M

Supply Voltage

Power Supply

Monitoring

-0.26V

-400V M

Supply Voltage

Power Supply

Monitoring

VCOM

System Supply Common

Power Supply

Monitoring

HVCOM M

High Voltage Supply
Common

Power Supply

Monitoring

1.7V

HVTEMP M

Voltage Amp Heat Sink
Temp.

High Voltage Amplifier

Monitoring

2.0V

ITEMP M

Current Amp Heat Sink
Temp.

Current Amplifier

Monitoring

-2.2V

ICOMPL M

Current Output Stage
Drive Level

Current Amplifier

Monitoring

I CAL/I FB

Shunt Voltage

Current Amplifier

Calibration, AC I
Operation

AC/AC

Calibrator Output

High Voltage Sense

Calibration

HVSENSE

Sense Amp Output

High Voltage Sense

Calibration

DAC OUT

DAC Output

High Voltage Sense

Monitoring

0.08, 2.0V

FAN M

Fan is On or Off

Power Supply

Monitoring

Note 1: For single limit values, the limit is equal to, or smaller in magnitude to, the value shown.