Introduction and Specifications

AC Voltage Specifications

1

1-15

Table 1-8. AC Voltage Secondary Performance and Operating Characteristics (continued)

5725A Amplifier:

Stability

±

1 °C [Note 1]

Temperature Coefficient

Adder

Distortion

Bandwidth

Range

Frequency

24 Hours

10°-40 °C

0°-10 °C

and

40°-50 °C

Load Regulation

[Note 2]

10 Hz-10 MHz

±

(% output)

Hz

±

(ppm output + mV)

±

(ppm output)/°C

±

(ppm output + mV)

150 pF

1000 pF

1100 V

0

40 - 1 k

0

1 k - 20 k

20 k - 50 k

50 k - 100 k

0

10 + .5

0

15 + 2

0

40 + 2

130 + 2

5

5

10

30

5

5

10

30

0

10 + 1

0

90 + 6

275 + 11

500 + 30

0.10

0.10

0.30

0.40

0.10

0.15

0.30

0.40

Voltage Range

Maximum Current Limits

Load Limits

Output display formats: Voltage or dBm,

dBm reference 600

.

2.2 V [Note 3]

22 V

220 V

50 mA, 0 °C-40 °C

20 mA, 40 °C-50 °C

>50

,

1000 pF

Minimum output:: 10 % on each range

External sense: Selectable for 2.2 V, 22 V, 220 V, and
1100 V ranges; 5700A/5720A <100 kHz, 5725A <30 kHz

1100 V

6 mA

600 pF

Settling time to full accuracy:

5725A Amplifier:

1000 pF [Note 2]

Frequency (Hz)

Settling time (seconds)

0

40 Hz-5 kHz

50 mA

<20

7

1100 V

0

5 kHz-30 kHz

70 mA

300 pF

120-120 k

5

30 kHz-100 kHz

70 mA
[Note 4]

150 pF

>120 k

2

Notes:

1 Stability specifications are included in Absolute Uncertainty values

for the primary specifications.

2. The 5725A will drive up to 1000 pF of load capacitance.

Uncertainty specifications include loads to 300 pF and 150 pF as
shown under "Load Limits." For capacitances up to the maximum of
1000 pF, add "Load Regulation."

3. 2.2 V Range, 100 kHz-1.2 MHz only: uncertainty specifications

cover loads to 10 mA or 1000 pF. For higher loads, load regulation
is added.

4. Applies from 0 °C to 40 °C

1000

100

10

0

3.5 V

10000

50 Hz

10 Hz

1 kHz

100 kHz

1 MHz

30 MHz

Voltage

Frequency

40 Hz

30 Hz

5725A

5700A-03

2.2 x 10

7

V-Hz

15 Hz

5700A/5720A

Volt-Hertz Capability

ahp160f.eps

Figure 1-1. Volt-Hertz Capability

+ 1 second for amplitude or frequency range change;
+ 2 seconds for 5700A/5720A 1100 V range;
+ 4 seconds for 5725A 1100 V range

Overshoot:: <10%
Common mode rejection: 140 dB, DC to 400 Hz

Frequency:

Ranges (Hz):

10.000-119.99
0.1200 k-1.1999 k, 1.200 k-11.999 k
12.00 k-119.99 k,120.0 k-1.1999

Uncertainty:

±

0.01 %

Resolution: 11.999 counts

Phase lock: Selectable rear panel BNC input
Phase uncertainty (except 1100 V range):

>30 Hz:

±

1

°

+ 0.05

°

/kHz), <30 Hz:

±

3

°

Input voltage: 1 V to 10 V rms sine wave (do not exceed
1 V for mV ranges)

Frequency range: 10 Hz to 1.1999 MHz

Lock range:

±

2 % of frequency

Lock-in time: Larger of 10/frequency or 10 msec

Phase reference: Selectable, rear panel BNC output
Range:

±

180

°

Phase Uncertainty (except 1100 V range):

±

1

°

at quadrature points (0

°

,

±

90

°

,

±

180

°

) elsewhere

±

2

°

Stability:

±

0.1

°

Resolution: 1

°

Output level: 2.5 V rms

±

0.2 V

Frequency range:

50 kHz to 1 kHz, usable 10 Hz to 1.1999 MHz