Introduction and Specifications

DC Voltage Specifications

1

1-9

Table 1-3. DC Voltage Secondary Performance Specifications and Operating Characteristics

Temperature Coefficient

Adder [Note 2]

Noise

Range

Stability [Note 1]

± 1 °C

10 °-40 °C

0 °-10 °C

and

Linearity

± 1 °C

Bandwidth

0.1-10 Hz

Bandwidth

10 Hz-10 kHz

24 Hours

40°-50 °C

pk-pk

RMS

±

(ppm output +

µ

V)

±

(ppm output +

µ

V)/°C

±

(ppm output +

µ

V)

µ

V

220 mV

2.2 V

11 V

22 V

220 V

1100 V

0.3 + 0.3

0.3 + 1

0.3 + 2.5

0.4 + 5

0.5 + 40

0.5 + 200

0

0.4 + 0.1

0

0.3 + 0.1

0.15 + 0.2

0

0.2 + 0.4

0

0.3 + 5

0

0.5 + 10

1.5 + 0.5

1.5 + 2

0

1 + 1.5

1.5 + 3

1.5 + 40

0

3 + 200

0

1 + 0.2

0

1 + 0.6

0.3 + 2

0.3 + 4

0

1 + 40

0

1 + 200

0.15 + 0.1

0.15 + 0.4

0.15 + 2

0.15 + 4

0.15 + 60

0.15 + 300

5

15

50

50

150

500

Notes:

1. Stability specifications are included in the Absolute Uncertainty values in the primary specification

tables.

2. Temperature coefficient is an adder to uncertainty specifications that does

not apply unless operating

more than

±

5 °C from calibration temperature.

Minimum output: 0 V for all ranges, except 100 V for 1100 V range

Maximum load: 50 mA for 2.2 V through 220 V ranges; 20 mA for 1100 V range; 50

output impedance

on 220 mV range; all ranges <1000 pF, >25

Load regulation: <(0.2 ppm of output + 0.1ppm of range), full load to no load

Line regulation: <0.1 ppm change,

±

10 % of selected nominal line

Settling time: 3 seconds to full accuracy; + 1 second for range or polarity change; + 1 second for 1100 V
range

Overshoot: <5 %

Common mode rejection: 140 dB, DC to 400 Hz

Remote sensing: Available 0 V to

±

1100 V, on 2.2 V through 1100 V ranges