5725a – Fluke 5725A User Manual

Page 40

Advertising
background image

5725A

Instruction Manual

1-24

Table 1-17. 5720A Series II AC Current Specifications: 95 % Confidence Level

5720A

95%

95 % Confidence Level

Range

Resolution

Frequency

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >0.4 V/m but

3 V/m,

add 1 % of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

Hz

±

(ppm output + nA)

±

(ppm output + nA)

220

µ

A

1 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

210 + 16

0

130 + 10

0

100 + 8

0

240 + 12

0

800 + 65

0

230 + 16

0

140 + 10

0

110 + 8

0

250 + 12

0

900 + 65

0

240 + 16

0

150 + 10

0

115 + 8

0

270 + 12

1000 + 65

0

250 + 16

0

160 + 10

0

120 + 8

0

280 + 12

1100 + 65

0

210 + 16

0

110 + 10

00

80 + 8

0

200 + 12

0

700 + 65

0

230 + 16

0

130 + 10

00

90 + 8

0

230 + 12

0

800 + 65

2.2 mA

10 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

210 + 40

0

140 + 35

0

100 + 35

0

170 + 110

0

800 + 650

0

230 + 40

0

140 + 35

0

110 + 35

0

180 + 110

0

900 + 650

0

240 + 40

0

150 + 35

0

115 + 35

0

190 + 110

1000 + 650

0

250 + 40

0

160 + 35

0

120 + 35

0

200 + 110

1100 + 650

0

210 + 40

0

110 + 35

00

80 + 35

0

200 + 110

0

700 + 650

0

230 + 40

0

130 + 35

00

90 + 35

0

230 + 110

0

800 + 650

22 mA

100 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

210 + 400

0

130 + 350

0

100 + 350

0

170 + 550

0

800 + 5000

0

230 + 400

0

140 + 350

0

110 + 350

0

180 + 550

0

900 + 5000

0

240 + 400

0

150 + 350

0

115 + 350

0

190 + 550

1000 + 5000

0

250 + 400

0

160 + 350

0

120 + 350

0

200 + 550

1100 + 5000

0

210 + 400

0

110 + 350

00

80 + 350

0

200 + 550

0

700 + 5000

0

230 + 400

0

130 + 350

00

90 + 350

0

230 + 550

0

800 + 5000

Hz

±

(ppm output +

µ

A)

±

(ppm output +

µ

A)

220 mA

1

µ

A

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

210 + 4

0

130 + 3.5

0

100 + 2.5

0

170 + 3.5

0

800 + 10

0

230 + 4

0

140 + 3.5

0

110 + 2.5

0

180 + 3.5

0

900 + 10

0

240 + 4

0

150 + 3.5

0

115 + 2.5

0

190 + 3.5

1000 + 10

0

250 + 4

0

160 + 3.5

0

120 + 2.5

0

200 + 3.5

1100 + 10

0

210 + 4

0

110 + 3.5

00

80 + 2.5

0

200 + 3.5

0

700 + 10

0

230 + 4

0

130 + 3.5

0

90 + 2.5

0

230 + 3.5

0

800 + 10

2.2 A

10

µ

A

0

20 - 1 k

0

1 k - 5 k

0

5 k - 10 k

0

230 + 35

0

350 + 80

5000 + 160

0

240 + 35

0

390 + 80

6000 + 160

0

250 + 35

0

420 + 80

6500 + 160

0

260 + 35

0

450 + 80

7000 + 160

0

250 + 35

0

400 + 80

5000 + 160

0

300 + 35

0

440 + 80

6000 + 160

5725A Amplifier:

11 A

100

µ

A

0

40 - 1 k

0

1 k - 5 k

0

5 k - 10 k

0

370 + 170

0

800 + 380

3000 + 750

0

400 + 170

0

850 + 380

3300 + 750

0

440 + 170

0

900 + 380

3500 + 750

0

460 + 170

0

950 + 380

3600 + 750

0

300 + 170

0

700 + 380

2800 + 750

0

330 + 170

0

800 + 380

3200 + 750

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220

µ

A and

2.2 mA ranges are increased by 1.3 x plus 2

µ

A when supplied through 5725A terminals. Specifications are

otherwise identical for all output locations.