Figure 6-4. Initial Access Procedure

6.

Verify that both LEDs CR317 and CR318 on the Power Supply assembly (A4) are
off. If either is lit after three minutes, there is a fault in the high voltage capacitor
bleed-off path.

Warning

Do not rely on an off condition as an indication that dangerous
voltage no longer exists on the filter capacitors. Always use a
voltmeter to check for high voltage between TP307 and TP301,
and between TP307 and TP304.

7.

To confirm that high voltage has dissipated, set a DMM to the 1000V dc range, and
take a reading with the common lead to TP307 and the high lead to TP301 on the
Power Supply assembly. This is the +400V filter capacitor test point, and it should
be at a safe level near zero.