Fluke 5725A User Manual

Page 130

Advertising
background image

5725A

Instruction Manual

6-16

J222

FRONT OUTPUT

CABLE

REAR OUTPUT

CABLE

CURRENT

AMPLIFIER

(A2)

SECURE UNUSED

CABLE

aq23f.eps

Figure 6-7. Enabling Front or Rear Binding Posts

Caution

Make sure that the unused connector is neatly secured in the
recessed area of the chassis side, away from the High Voltage
Sense assembly as Figure 6-7 shows. If it is allowed to hang
down, shorting can occur.

8.

Reconfiguration is complete. Reinstall the Current Amplifier and High Voltage
Amplifier assemblies, being careful to correctly seat each assembly into the
connector on the Interconnect assembly.

9.

Reinstall the top cover.