3-10. Voltage Operate

CURRENT

OUTPUT

HI

LO

VOLTAGE MODE

CURRENT MODE

5725A STANDBY

FAULT

OVERLOAD

CABLE OFF

11A

MAX

2OV

PK

MAX

NO OUTPUT

HI

HI

LO

LO

AUX

CURRENT

GUARD

GROUND

WIDEBAND

SENSE

V A

OUTPUT

V A

µ

VOLTAGE

OUTPUT

HERE

HI

BOOST

LIT

aq11i.eps

Figure 3-6. Voltage Operate Mode

Warning

Both the 5700A and 5725A can produce fatal voltages. boosted
voltage operation produces high voltage at higher current
levels than normally available from the 5700A. during voltage
operation, the potential risk of injury or fatal accident is greater
than during normal operation.

Voltage operate mode is when only the VOLTAGE MODE indicator is lit. In voltage
operate mode, amplified voltage output is available at the 5700A binding posts. Voltage
operate occurs if "5725A" is selected in a 5700A setup menu as the voltage amplifier,
and the next two conditions are true:

An ac voltage 220V or above is showing on the 5700A Output Display with the
OPERATE annunciator lit.

The BOOST indicator on the 5700A is lit.

Note

If condition 1 is true, the BOOST indicator can be toggled on and off by
pressing the 5700A BOOST key. If the BOOST indicator is off, then the
5725A is in standby mode.

Note

Under certain settings of ac voltage, especially high voltage between 1 to 5
kHz, a whine from the 5725A is audible. This originates from the output
transformer, and is normal.