5725a – Fluke 5725A User Manual

Page 42

Advertising
background image

5725A

Instruction Manual

1-26

Table 1-19. 5700A Series II AC Current Specifications: 95 % Confidence Level

5700A

95%

95 % Confidence Level

Range

Resolution Frequency

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >0.4 V/m but

3 V/m,

add 1 % of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

Hz

±

(ppm output + nA)

±

(ppm output + nA)

220

µ

A

1 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

550 + 25

0

280 + 20

0

100 + 16

0

400 + 40

1300 + 80

0

600 + 25

0

310 + 20

0

120 + 16

0

500 + 40

1400 + 80

0

650 + 25

0

330 + 20

0

130 + 16

0

550 + 40

1500 + 80

0

700 + 25

0

350 + 20

0

140 + 16

0

600 + 40

1600 + 80

0

375 + 25

0

220 + 20

0

90 + 16

0

375 + 40

1200 + 80

0

400 + 25

0

250 + 20

0

100 + 16

0

400 + 40

1200 +80

2.2 mA

10 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

550 + 40

0

280 + 35

0

100 + 35

0

400 + 400

1300 + 800

0

600 + 40

0

310 + 35

0

120 + 35

0

500 + 400

1400 + 800

0

650 + 40

0

330 + 35

0

130 + 35

0

550 + 400

1500 + 800

0

700 + 40

0

350 + 35

0

140 + 35

0

600 + 400

1600 + 800

0

375 + 40

0

220 + 35

00

90 + 35

0

375 + 400

1200 + 800

0

400 + 40

0

250 + 35

0

100 + 35

0

400 + 400

1200 + 800

22 mA

100 nA

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

550 + 400

0

280 + 350

0

100 + 350

0

400 + 4000

1300 + 8000

0

600 + 400

0

310 + 350

0

120 + 350

0

500 + 4000

1400 + 8000

0

650 + 400

0

330 + 350

0

130 + 350

0

550 + 4000

1500 + 8000

0

700 + 400

0

350 + 350

0

140 + 350

0

600 + 4000

1600 + 8000

0

375 + 400

0

220 + 350

00

90 + 350

0

375 + 4000

1200 + 8000

0

400 + 400

0

250 + 350

0

100 + 350

0

400 + 4000

1200 + 8000

Hz

±

(ppm output +

µ

A)

±

(ppm output +

µ

A)

220 mA

1

µ

A

0

10 - 20

0

20 - 40

0

40 - 1 k

0

1k - 5 k

0

5k - 10 k

0

550 + 4

0

280 + 3.5

0

100 + 3.5

0

400 + 40

1300 + 80

0

600 + 4

0

310 + 3.5

0

120 + 3.5

0

500 + 40

1400 + 80

0

650 + 4

0

330 + 3.5

0

130 + 3.5

0

550 + 40

1500 + 80

0

700 + 4

0

350 + 3.5

0

140 + 3.5

0

600 + 40

1600 + 80

0

375 + 4

0

220 + 3.5

00

90 + 3.5

0

375 + 40

1200 + 80

0

400 + 4

0

250 + 3.5

0

100 + 3.5

0

400 + 40

1200 + 80

2.2 A

10

µ

A

0

20 - 1 k

0

1 k - 5 k

0

5 k - 10 k

0

500 + 35

0

600 + 80

6500 + 160

0

550 + 35

0

650 + 80

7500 + 160

0

600 + 35

0

700 + 80

8000 + 1600

0

650 + 35

0

750 + 80

8500 + 160

0

500 + 35

0

550 + 80

6000 + 160

0

550 + 35

0

650 + 80

7000 + 160

5725A Amplifier:

11 A

100

µ

A

0

40 - 1 k

0

1 k - 5 k

0

5 k - 10 k

0

370 + 170

0

800 + 380

3000 + 750

0

400 + 170

0

850 + 380

3300 + 750

0

440 + 170

0

900 + 380

3500 + 750

0

460 + 170

0

950 + 380

3600 + 750

0

300 + 170

0

700 + 380

2800 + 750

0

330 + 170

0

800 + 380

3200 + 750

Note: Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220

µ

A and

2.2 mA ranges are increased by a factor of 1.3 plus 2

µ

A when supplied through 5725A terminals.

Specifications are otherwise identical for all output locations.