Selecting Line Voltage

2

2-7

2-7. Selecting Line Voltage

The 5725A arrives from the factory configured for the line voltage normally appropriate
for the country of purchase, or as specified at the time of purchase. Check the line power
label on the rear panel of the 5725A to verify that the line voltage matches local line
power. Figure 2-3 shows the location of the line power label and switches, and the
switch settings for each line voltage.

The 5725A can be set to operate from eight different nominal line voltages; each voltage
setting has a voltage tolerance of

±

10% and a frequency range of 47 to 63 Hz. The line

voltage switches are located on the bottom left side of the rear panel.

To change the line voltage setting, set the line voltage selection switches to the correct
setting shown in Figure 2-3.

100V
110V
115V

120V

S2 S3 S4

VOLTAGE

SELECTION

FUSE

F8A 250V

(FAST)

200V
220V
230V
240V

S2 S3 S4

F4A 250V

(FAST)

JOHN FLUKE MFG. CO., INC.

MADE IN U.S.A.

PATENTS PENDING

NO INTERNAL USER SERVICEABLE
PARTS REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL

TO CLEAN FILTER
REMOVE FROM INSTRUMENT
AND FLUSH WITH WARM
SOAPY WATER

5700A CALIBRATOR

20V PK

MAX

CHASSIS
GROUND

FUSE

WARNING

GROUNDING
CONNECTOR IN POWER CORD
MUST BE CONNECTED TO
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.

47-63 Hz
750VA MAX

CAUTION

FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.

S4

S3

S2

100V
110V
115V

120V

S2 S3 S4

VOLTAGE

SELECTION

FUSE

F8A 250V

(FAST)

200V
220V
230V
240V

S2 S3 S4

F4A 250V

(FAST)

CAUTION

FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.

11A MAX

HI

LO

CALIBRATION

ENABLE NORMAL

CURRENT

OUTPUT

!

!

!

!

!

aq4f.eps

Figure 2-3. Line Power Label and Switch Location