19. auxiliary amplifier specifications – Fluke 5725A User Manual

Page 46

Advertising
background image

5725A

Instruction Manual

1-30

1-19. Auxiliary Amplifier Specifications

For complete specifications, see the 5205A and 5220A Operators Manuals.

5205A (220V - 1100 V ac, 0 V - 1100 V dc)

Overshoot: < 10 %

Distortion (bandwidth 10 Hz - 1 MHz):

10 Hz - 20 kHz

20 kHz - 50 kHz

50 kHz - 100 kHz

0.07 %

0.2 %

0.25 %

Frequency

(Hz)

90 Day Accuracy

at 23°

±

5 °C

±

(% output + % range)

Temperature Coefficient for

0°-18 °C and 28°-50 °C

±

(ppm output + ppm range)/ °C

00

dc

0

10 - 40

0

40 - 20 k

20 k - 50 k

50 k - 100 k

0.05 + 0.005

0.15 + 0.005

0.04 + 0.004

0.08 + 0.006

0

0.1 + 0.01

15 + 3

45 + 3

15 + 3

50 + 10

70 + 20

5220A (AC Current, 180-day specifications):

Accuracy:

20 Hz - 1 kHz
1 kHz - 5 kHz

0.07% + 1 mA
(0.07% + 1mA) x frequency in kHz

Temperature Coefficient (0° - 18 °C and 28° - 50 °C): (0.003 % + 100

µ

A)/°C

Distortion (bandwidth 300 kHz):

20 Hz - 1 kHz
1 kHz - 5 kHz

0.1% + 1 mA
(0.1% + 1 mA) x frequency in kHz

Note: 5700A/5720A combined with 5220A is not specified for inductive loads.