Troubleshooting

Self Diagnostics

7

7-5

Table 7-1. 5725A Fault Codes (cont)

FAULT

NO.

MESSAGE

EXPLANATION

228

5725 Software TRAP

229

5725 Cable Was Off

230

5725 RESET (power-up or watchdog timer)

231

5725 Guard-Crossing Timeout

232

5725 Illegal/Unexecutable Command

233

5725 Non-Maskable Interrupt Occurred

Either internal fault or external transient condition

234

5725 HVCLEAR Circuit Activated

Internal fault, requires service

235

Output Tripped To Standby

HVCLEAR circuit activated most likely because of
external transient condition

841

No 5725 Available For Selected Output

850

Can’t Store, 5725 Switch In NORMAL

851

Can’t Format, 5725 Switch In NORMAL

3202

5725 ACV Sense Amp Fault

Requires service

3203

5725 ACV Standby 5725 Fault

Requires service

3204

5725 ACV Operate Wouldn’t Converge

Requires service

3205

5725 ACV Operate 5725 Fault

Requires service

3206

5725 ACV Cal Sensor Test Died

Requires service

3207

5725 ACV Cal Sensor Fault

Requires service

3208

5725 Current Path To Shunt Open

Requires service

3209

5725 Shunt Sense Open

Requires service

3210

5725 Shunt Measurement Out Of
Tolerance

Requires service

3211

5725 Current Amplifier Offset Too Large

Requires service

3212

5725 Current Drive Path Open

Requires service

3213

5725 Current Error Amplifier Failure

Requires service

6.

To test the current function, press the "Current" softkey. A sequence of self tests is
run. A brief description of each test follows:

a.

5725A Current Shunt: Checks the current cabling and the calibration amplifier
on the Current Amplifier assembly (A2).

b.

5725A Current Zero: Grounds the input and shorts the output of the Current
Amplifier, and verifies that the output is close to 0A.

c.

5725A Current Gain: Shorts the output of the Current Amplifier, applies a
voltage input, and verifies the output current. This confirms integrity of the error
amplifier circuit in the Current Amplifier assembly.

d.

5725A Current Monitor: Runs the analog monitor circuit on the High Voltage
Sense Assembly (A6) in the current function and verifies that it operates
correctly.