Fluke 5725A User Manual

Page 44

Advertising
background image

5725A

Instruction Manual

1-28

1-17. Wideband AC Voltage (Option 5700-03) Specifications

Table 1-21. Wideband AC Voltage (Option 5700-03) Specifications

Specifications apply to the end of the cable and 50

termination used for calibration

Range

Resolution

Absolute Uncertainty

±

5 °C from calibration temperature

30 Hz-500 kHz

Volts

dBm

24 Hours

90 Days

180 Days

1 Year

±

(% output +

µ

V)

1.1 mV

-46

10 nV

0.4 + 0.4

0

0.5 + 0.4

0

0.6 + 0.4

0.8 + 2

3 mV

-37

10 nV

0.4 + 1

0.45 + 1

0

0.5 + 1

0.7 + 3

11 mV

-26

100 nV

0.2 + 4

0.35 + 4

0

0.5 + 4

0.7 + 8

33 mV

-17

100 nV

0.2 + 10

0

0.3 + 10

0.45 + 10

0.6 + 16

110 mV

-6.2

1

µ

V

0.2 + 40

0

0.3 + 40

0.45 + 40

0.6 + 40

330 mV

+3.4

1

µ

V

0.2 + 100

0

2.5 + 100

0.35 + 100

0.5 + 100

1.1 V

+14

10

µ

V

0.2 + 400

0.25 + 400

0.35 + 400

0.5 + 400

3.5 V

+24

10

µ

V

0.15 + 500

0

0.2 + 500

0

0.3 + 500

0.4 + 500

Frequency

Frequency

Resolution

Amplitude Flatness, 1 kHz Reference

Voltage Range

Temperature

Coefficient

Settling

Time To

Full

Accuracy

Harmonic

Distortion

1.1 mV

3 mV

>3 mV

Hz

Hz

±

(% output + floor indicated)

±

ppm/°C

Seconds

dB

00

10 - 30

0.01

0.3

0.3

0.3

100

7

-40

00

30 - 120

0.01

0.1

0.1

0.1

100

7

-40

0

120 - 1.2 k

0.1

0.1

0.1

0.1

100

5

-40

0

1.2 k - 12 k

1

0.1

0.1

0.1

100

5

-40

0

12 k - 120 k

10

0.1

0.1

0.1

100

5

-40

120 k - 1.2 M

100

0

0.2 + 3

µ

V

0

0.1 + 3

µ

V

0.1 + 3

µ

V

100

5

-40

1.2 M - 2 M

100 k

0

0.2 + 3

µ

V

0

0.1 + 3

µ

V

0.1 + 3

µ

V

100

0.5

-40

0

2 M - 10 M

100 k

0

0.4 + 3

µ

V

0

0.3 + 3

µ

V

0.2 + 3

µ

V

100

0.5

-40

10 M - 20 M

1 M

0

0.6 + 3

µ

V

0

0.5 + 3

µ

V

0.4 + 3

µ

V

150

0.5

-34

20 M - 30 M

1 M

10.5 + 15

µ

V

10.5 + 3

µ

V

0

1 + 3

µ

V

300

0.5

-34

Additional Operating Information:

dBm reference = 50

Range boundaries are at voltage points, dBm levels are approximate.

dBm = 10 log

(

Power

1 mW

)

; 0.22361 V across 50

= 1 mW or 0 dBm

Minimum output: 300

µ

V (-57 dBm)

Frequency uncertainty:

±

0.01 %

Frequency resolution: 11,999 counts to 1.1999 MHz, 119 counts to 30 MHz.

Overload protection: A short circuit on the wideband output will not result in damage. After settling time, normal operation is

restored upon removal.