Fluke 5725A User Manual

Page 37

Advertising
background image

Introduction and Specifications

DC Current Specifications

1

1-21

Table 1-14. 5700A Series II DC Current Specifications: 99 % and 95 % Confidence Levels

5700A

99%

99 % Confidence Level

Range

Resolution

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >0.4 V/m but

3 V/m,

add 1% of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

nA

±

(ppm output + nA)

±

(ppm output + nA)

220

µ

A

2.2 mA

22 mA

0.1

1

10

0

45 + 10

0

45 + 10

0

45 + 100

0

50 + 10

0

50 + 10

0

50 + 100

0

55 + 10

0

55 + 10

0

55 + 100

0

60 + 10

0

60 + 10

0

60 + 100

0

24 + 2

0

24 + 5

0

24 + 50

0

26 + 2

0

26 + 5

0

26 + 50

µ

A

±

(ppm output +

µ

A)

±

(ppm output +

µ

A)

220 mA

2.2 A

[Note 1]

0.1

1

0

55 + 1

0

75 + 30

0

60 + 1

0

80 + 30

0

65 + 1

0

90 + 30

0

70 + 1

0

95 + 30

0

26 + .3

0

40 + 7

0

30 + .3

0

45 + 7

5725A Amplifier:

11 A

10

330 + 470

340 + 480

350 + 480

360 + 480

100 + 130

110 + 130

5700A

95%

95 % Confidence Level

Range

Resolution

Absolute Uncertainty

±

5 °C from calibration temperature

For fields strengths >0.4 V/m but

3 V/m,

add 1 % of range.

Relative Uncertainty

±

1 °C

24 Hours

90 Days

180 Days

1 Year

24 Hours

90 Days

nA

±

(ppm output + nA)

±

(ppm output + nA)

220

µ

A

2.2 mA

22 mA

0.1

1

10

0

35 + 8

0

35 + 8

0

35 + 80

0

40 + 8

0

40 + 8

0

40 + 80

0

45 + 8

0

45 + 8

0

45 + 80

0

50 + 8

0

50 + 8

0

50 + 80

0

20 + 1.6

0

20 + 4

0

20 + 80

0

22 + 1.6

0

22 + 4

0

22 + 80

µ

A

±

(ppm output +

µ

A)

±

(ppm output +

µ

A)

220 mA

2.2 A

[Note 1]

0.1

1

0

45 + 0.8

0

60 + 25

0

50 + 0.8

0

65 + 25

0

55 + 0.8

0

75 + 25

0

60 + 0.8

0

80 + 25

0

22 + 0.25

0

35 + 6

0

25 + 0.25

0

40 + 6

5725A Amplifier:

11 A

10

330 + 470

340 + 480

350 + 480

360 + 480

100 + 130

110 + 130

Note:

Maximum output from the calibrator’s terminals is 2.2 A. Uncertainty specifications for 220 mA and

2.2 mA ranges are increased by a factor of 1.3 when supplied through 5725A terminals.

Specifications are otherwise identical for all output locations.

1. Add to uncertainty specifications:

±

200 x

I

2

ppm for >100 mA on 220 mA range

±

10 x

I

2

ppm for >1A on 2.2A range