Fluke 5725A User Manual

Page 34

Advertising
background image

5725A

Instruction Manual

1-18

Table 1-11. Resistance Secondary Performance Specifications and Operating Characteristics

Nominal

Value

Stability

±

1 °C

[Note 1]

Temperature

Coefficient Adder

[Note 2]

Full Spec

Load Range

Maximum

Peak

Current

Maximum

Difference

of

Characterized

Two-Wire Adder

active

compensation

[Note 4]

10°-40 °C

0°-10 °C

and

[Note 3]

to Nominal

Value

Lead Resistance

24 Hours

40°-50 °C

0.1

1

±

ppm

±

ppm/°C

mA

mA

±

ppm

±

m

0

00

8 - 500

500

2 +

Im

4 V

µ

4 +

Im

4 V

µ

1

32

4

5

00

8 - 100

700

500

2 +

Im

4 V

µ

4 +

Im

4 V

µ

1.9

25

6

7

00

8 - 100

500

500

2 +

Im

4 V

µ

4 +

Im

4 V

µ

10

5

2

3

00

8 - 11

220

300

2 +

Im

4 V

µ

4 +

Im

4 V

µ

19

4

2

3

00

8 - 11

160

300

2 +

Im

4 V

µ

4 +

Im

4 V

µ

100

2

2

3

00

8 - 11

70

150

2 +

Im

4 V

µ

4 +

Im

4 V

µ

190

2

2

3

00

8 - 11

50

150

2 +

Im

4 V

µ

4 +

Im

4 V

µ

1 k

2

2

3

00

1 - 2

22

150

10

15

1.9 k

2

2

3

00

1 - 1.5

16

150

10

15

10 k

2

2

3

100 - 500

µ

A

7

150

50

60

19 k

2

2

3

0

50 - 250

µ

A

5

150

100

120

100 k

2

2

3

10 - 100

µ

A

1

150

I

m

= Current

produced by
Ohmmeter

190 k

2

2

3

00

5 - 50

µ

A

500

µ

A

150

1 M

2.5

2.5

6

00

5 - 20

µ

A

100

µ

A

200

1.9 M

3.5

3

10

0

2.5 - 10

µ

A

50

µ

A

200

10 M

10

5

20

0

0.5 - 2

µ

A

10

µ

A

300

19 M

20

8

40

0.25 - 1

µ

A

5

µ

A

300

100 M

50

12

100

0

50 - 200 nA

1

µ

A

500

Notes:

1. Stability specifications are included in the Absolute Uncertainty values in the primary specification tables.

2. Temperature coefficient is an adder to uncertainty specifications that does not apply unless operated

more than 5 °C from calibration temperature, or calibrated outside the range 19 °C to 24 °C. Two
examples:

Calibrate at 20 °C: Temperature coefficient adder is not required unless operated below 15 °C or
above 25 °C.

Calibrate at 26 °C: Add 2 °C temperature coefficient adder. Additional temperature coefficient adder is
not required unless operated below 21 °C or above 31 °C.

3. Refer to current derating factors table for loads outside of this range.

4. Active two-wire compensation may be selected for values less than 100 k

, with either the front panel or

the meter input terminals as reference plane. Active compensation is limited to 11 mA load, and to 2 V
burden. Two-wire compensation can be used only with

-meters that source continuous (not pulsed) dc

current.