5725A

Instruction Manual

6-12

CURRENT

AMPLIFIER

(A2)

BRIDGE

RECTIFIER
HEAT SINK

DRESS CABLES

AS SHOWN

aq18f.eps

Figure 6-6. Correct Way to Dress Cables

6-11. Accessing the High Voltage Amplifier (A3)

The High Voltage Amplifier assembly lays flat on the top of the instrument, toward the
front. To access the High Voltage Amplifier assembly, simply do the Initial Access
Procedure. This leaves the assembly in the service position.

Note

When reinstalling the assembly, lower it into place squarely to make sure
its connector seats correctly in the Interconnect assembly connector.