Fluke 5725A User Manual

Page 35

Advertising
background image

Introduction and Specifications

Resistance Specifications

1

1-19

Table 1-12. Current Derating Factors

Nominal Value

Value of Derating Factor K for Over or Under Current

Two-Wire Comp

I

<

I

L

[Note 1]

Four-Wire

I

<

I

L

[Note 1]

Four-Wire

I

U

<

I

<

I

max

[Note 2]

SHORT

1

1.9

10

19

100

190

1 k

1.9 k

10 k

19 k

100 k

190 k

1 M

1.9 M

10 M

19 M

100 M

4.4

4.4

4.4

4.4

4.4

4.4

4.4

4.4

4.4

5000

5000






0.3

300

160

30

16

3.5

2.5

0.4

0.4

50

50

7.5

4.0

1.0

0.53

0.2

0.53

0.1

4 x10

-5

1.5 x 10

-4

1.6 x 10

-3

3 x 10

-3

1 x 10

-2

1.9 x 10

-2

0.1

0.19

2.0

3.8

2 x 10

-5

3.8 x 10

-5

1.5 x 10

-4

2.9 x 10

-4

1 x 10

-3

1.9 x 10

-3

Notes:

1. For

I

<

I

L

,

errors occur due to thermally generated voltages w

i

th

i

n the 5720A. Use the follow

i

ng

equat

i

on to determ

i

ne the error, and add th

i

s error to the correspond

i

ng uncerta

i

nty or stab

i

l

i

ty

spec

i

f

i

cat

i

on.

Error = K(

I

L

-

I

)/(

I

L

x

I

)

Where: Error

i

s

i

n m

for all two-w

i

re comp values and four-w

i

re short, and

i

n ppm for the rema

i

n

i

ng

four-w

i

re values.

K

i

s the constant from the above table;

I

and

I

L

are expressed

i

n mA for short to 1.9 k

;

I

and

I

L

are expressed

i

n

µ

A for 10 k

to 100 M

2. For

I

U

<

I

<

I

MAX

errors occur due to self-heat

i

ng of the res

i

stors

i

n the cal

i

brator. Use the following

equat

i

on to determ

i

ne the error

i

n ppm and add th

i

s error to the correspond

i

ng uncerta

i

nty or stab

i

l

i

ty

spec

i

f

i

cat

i

on.

Error

i

n ppm = K(

I

2

-

I

U

2

)

Where: K

i

s the constant from the above table;

I

and

I

U

are expressed

i

n mA for short to 19 k

;

I

and

I

U

are expressed

i

n

µ

A for 100 k

to 100 M