Microtest TSB-67 - 568A Annex E User Manual

Page 54

Advertising
background image

Frequency range

PentaScanner+

: 1.0 to 100 MHz

PentaScanner

/

PentaScanner

Cable Admin

: 1.0 to 100 MHz

Sweep mode: Swept 1- 100 MHz using 1 MHz step or fixed points (1.0, 2.0, 4.0,5,0, 8.0,
10, 16.0, 20, 25, 31.25, 62.5, and 100 MHz)

Dynamic Range: 0 - 50 dB

Measurement Accuracy:

PentaScanner+

: Calculated per TIA TSB-67 error model (see NEXT and Attenuation

Specifications table)

PentaScanner

/

PentaScanner

Cable Admin

: 1.5 dB

Measurement resolution: 0.1 dB

Impedance

Range:

40 to 200 ohms

Accuracy:

± 5 ohms

Resolution: 2

ohms

Loop Resistance

Range:

0 to 10000 ohms

Accuracy:

± 0.3 ohms @ 0 to 15 ohms

± 2% @ 15 to 109 ohms (

PentaScanner+

)

± 2% @ 15 to 200 ohms (

PentaScanner

/

PentaScanner

Cable Admin

± 5% @ 109 to 5000 ohms (

PentaScanner+

)

± 5% @ 200 to 5000 ohms (

PentaScanner

/

PentaScanner

Cable

Admin

)

± 10% @ 5000 to 10000 ohms

Resolution:

0.1 ohms

0 to 200 ohms

1.0 ohms

200 to 9.999K ohms

Capacitance

Range:

0 to 100,000 pF

Accuracy:

± 20 pF

0 to 200 pF

± 10%

200 to 100,000 pF

Resolution:

5 pF

20 to 1000 pF

50 pF

1000 to 100,000 pF

Impulse Noise

Detects the occurrence of narrow impulse type events exceeding an adjustable threshold of

± 0.262, 0.524, 1.0, 2.0 or 2.5 volts. Instantly displays up to 200 impulse noise events per
second. Graphically logs and displays impulse events for up to 25 hours. Termination
impedance: 200 ohms.

Peak to Peak Noise

Measures the peak to peak voltage of steady state noise signals.

Range: 0 to 3,000 mVpp into 100 ohms.

Bandwidth: 3dB from 40 kHz to 30 MHz