Using multiple injectors, Injector calibration – Microtest TSB-67 - 568A Annex E User Manual

Page 8

Advertising
background image

Injector

Measurement

Required?

Description

Attenuation

Yes

The Injector generates the frequencies and the Scanner
measures the signal to determine the signal loss.

Autotest

Yes

The Injector is required for Attenuation, NEXT
measurements, Resistance and Wire Map during Autotest.

Capacitance

No

If connected, the Injector provides an “OPEN” for the
duration of the measurement.

Length

No

If connected, the injector provides an “OPEN” for the
duration of the measurement.

Impedance

No

If connected, the injector provides an “OPEN” for the
duration of the measurement.

NEXT - Scanner

Yes

The Injector provides a 100 ohm termination while the

End

Scanner measures NEXT.

NEXT - Injector

Yes

The Scanner provides a 100 ohm termination for the

End (2-Way

duration of the measurement.

Injector+/2-Way
Injector only)

Loop Resistance

Yes

If connected, the Injector is used to “SHORT” the wire pair
under test and the Scanner measures Loop Resistance. If the
2-Way Injector is not connected, the Scanner will display
“OVR” (Over Range).

Wire Map (Find

Yes

The Scanner communicates with the Injector across the link

Cable Faults)

under test to determine the wire connection from the
Scanner end to the Injector end.

Noise

No

The Injector is not required to measure impulse noise or
peak-to-peak noise.

Note: The Injector is not used when testing coaxial cable. See Appendix D: Working with
Coaxial Cable

for more information.

Using Multiple Injectors

One Scanner can store calibration data for up to five Injectors. To determine which Injector is
attached to the Scanner, press the

Extended Functions

key, then select

Injector Info

. The

Scanner will display the serial number and other information for the attached Injector.

Injector Calibration

Calibration of the Injector’s signal amplitude is required for accurate Attenuation measurements
and requires user intervention. To maintain accurate Attenuation measurements, the Injector must